Blood plasma chemistry protocol
| Prepared by: | MAM |
| Date Prepared: | April 6, 2001 |
| Code: | MP-004 |
| Reviewed by: | KLS |
| Revision #: | 4.0 |
| Date Revised: | 7/31/02 |
Calibrated commercial controls are run daily and must be within established ranges prior to analyzing samples.
| 0.0 | Abstract: The Beckman Coulter Synchron CX5 ("Chemistry Analyzer") is used to measure four blood plasma constituents for this PGA: HDL cholesterol (HDL), total cholesterol (CHOL), triglycerides (TG), and glucose (GLU). | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 1.0 | Instrument layout: A dedicated DOS-based desktop computer controls the programming of the Chemistry Analyzer. A dedicated printer prints the results as they are measured, and an electronic file is simultaneously transferred to a second, Windows-based computer, which stores the data files. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 2.0 | Reagents and expendables. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 2.1 Expendables: The following supplies and quantities are used for each sample tested (for all four of the above chemistries): 1.5µl Eppendorf tube (1), 0.5ml Beckman Coulter MicrotubeTM Tubecup ("sector cup") (2), and 200µl pipette tips (3-4, depending on whether dilution is necessary). | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
2.2 Reagents: The Chemistry Analyzer (or "CX5") uses Beckman Coulter three-compartment reagent cartridges for HDLc, CHOL, TG, and GLU. Each cartridge contains enough reagent for 300 tests (approx. 104mL). In addition, in order to run the HDL Cholesterol test, HDL Cholesterol Separation Reagent (15µl per sample) is needed. The bottle from Beckman Coulter contains a volume of 34ml. If the dilution of plasma samples becomes necessary due to low plasma volume, use 0.9% saline solution for the dilution.
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 2.3 Calibration reagents: The two calibration reagents are "Synchron® Systems HDL Cholesterol Calibrator" (for HDL only), and "Synchron® Systems Multi Calibrator" (for CHOL, GLU, and TG). | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 2.4 Controls: The controls for HDLc are Beckman CoulterTM VigilTM Lipid Control 1 and Beckman CoulterTM VigilTM Lipid Control 2. The controls for CHOL, GLU and TG are Synchron® Control Comprehensive Chemistry Control Serum Level 1, Level 2 and Level 3. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 3.0 | Setup: | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 3.1 Reagent preparation: Using the DOS computer, check the number of runs available for each reagent cartridge using the following sequence: from the Main Menu, press F2 ("Reagent Load"), and then F5 ("Reagent Status"). Next, page down through the reagents to check the number of runs left in each cartridge (under "Avail") for the four desired chemistries. Change the reagent cartridges as necessary, using the following procedure: from the Main Menu, press F2 ("Reagent Load"), and then page down through the reagents until the cursor is on the reagent that needs changing. Press Select to highlight it, and then press F1 ("Auto Load"). When prompted to do so by the CX5, open the reagent door and remove the empty cartridge, taking care to slide it past the laser bar code reader (listen for the "beep"). Obtain the new reagent cartridge in the adjacent refrigerator, and load it past the bar code reader in the same manner into the CX5 when prompted to do so, after which the CX5 will check the levels in the new cartridge. After loading a new reagent cartridge, you must calibrate the CX5 for that reagent. Refer to calibration procedures below. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 3.2 Calibration: From the Main Menu, press F3 ("Cal"), and then page down through the reagents until the cursor is over the reagent that needs to be calibrated. Press Select to highlight it, and then press F1 ("Cal Cup Assignment"). Enter the number of an available sector when prompted, and press Enter. Place a single 0.5 ml sector cup in the first position ("cell") of the previously assigned sector. Using the appropriate calibration reagent (described above in Section 2.3), squeeze several drops from the bottle into the sector cup, place the sector into the CX5, and press the green Start button on the keyboard. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 3.3 Control preparation: Controls are run using the same procedure used for samples described below. The Lipid Controls (Levels 1 and 2) used for HDL must be reacted with the HDL Cholesterol Separation Reagent, spun down, and extracted in the same manner, whereas the Comprehensive Chemistry Control Serum Levels 1, 2 and 3 are used directly. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 3.4 Sample preparation: The chemistry tests are run on mouse plasma which has been extracted from mouse blood in the laboratory. The mice are fasted for four hours, beginning one hour after the start of their light cycle, prior to the collection of the blood for the chemistry tests. At the conclusion of the fast, approximately 225µl of blood (6-8 drops plus one full capillary tube) are collected from each mouse (depending on its size) using a retro-orbital bleed. The blood is collected into previously labeled 1.5 ml Eppendorf tubes containing 7.5µl of Heparin 1000u/ml and put on ice immediately. It is then spun down in a microcentrifuge at 14,000 rpm for 5 minutes. A minimum of 100µl of translucent blood plasma is collected from the upper layer using a 200µl Pipetman, leaving the blood cells behind in the Eppendorf tube which is discarded. This plasma is pipetted into a previously labeled 0.5ml Eppendorf tube and frozen for future analysis. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
3.5 HDL cholesterol: [Procedure used by the Heart, Lung, Blood & Sleep Disorders Center at The Jackson Laboratory differs slightly from the Beckman protocol]. Note-The following procedure uses six sectors (42 samples) as an example run which is a convenient size to run at a single time. If more or fewer samples need to be run, modify the procedure appropriately.
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 3.6 | Cholesterol, glucose and triglycerides:
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.0 | Run: | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.1 The four chemistry tests used in the HLB PGA protocol cannot be run together and must be divided into two groups. The HDL test must be run separately because it utilizes a reagent to separate the two types of cholesterol from one another. The CHOL, GLU, and Tg tests can be run at the same time. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.2 The tests are run in the following order: 1) HDL, 2) CHOL & GLU (labelled as a CHOL level), 3) TG. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.3 The plasma used for the HDL test cannot be diluted but is calculated as a dilution factor of 2 because of the 35µl PEG and 35µl plasma.This should be removed first. If there is not enough plasma to run the other three tests, the plasma may be diluted in a 1:1 or 1:2 ratio. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.4 Using the DOS computer, be sure there are enough runs left in the reagent cartridge for the number of samples to be tested. Follow the procedure described in Section 3.1, Reagent Preparation. This is maintained by the Diagnostic Lab Staff. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.5 Program the CX5 to test the samples using the following procedure. From the Main Menu, press F1 ("Sample Program"), then F2 ("Program Batch"). Next, enter the sector numbers to program (e.g. "1-6") where prompted, and press Enter. The computer will respond with "Batch mode activated, (42) cups possible." Enter number of cups in batch (e.g. "42") where prompted, and press Enter. Select the desired chemistries by scrolling through the list of possibilities, and press Select appropriately. Press F8 ("End Batch") to program the CX5. "Please Wait -- Writing to Disk" will be displayed as the information is recorded. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.6 If the status of the CX5 is "Waiting," press the green Start button on the keyboard to begin testing the samples. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.7 If the CX5 is already operating, press the "Load" button on the CX5 itself to load the samples. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.8 Load three of the programmed sectors into the CX5 and wait while they are automatically loaded onto the "Sample Wheel." Load new sectors onto the wheel as space becomes available and remove sectors from the machine as the testing of those sectors is completed. It is important to monitor the progress of the CX5 closely in order to promptly troubleshoot any problems arising. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.9 Clear the sectors from the computer so that they may be re-programmed for future tests. From the Main Menu, press F1 ("Sample Program"), then F5 ("Clear Sectors"), and then enter the sector numbers to clear (e.g. "1-6") when prompted. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 4.10 Remove and discard the sector cups once all desired tests have been run. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 5.0 | Clean Up: Discard the Eppendorf tubes, pipette tips, sector cups, and reagent cartridges into biohazard waste containers. Clean up any spilled liquids. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 6.0 | Data Reduction: | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 6.1 The results will be printed on the dedicated dot-matrix printer as the CX5 finishes the test(s) per sample. Carefully monitor the activity of the printer to ensure uninterrupted feeding of the continuous-feed paper, and to collect the printed results in an accordion-folded stack, which will prevent the output paper from getting dragged into the printer with the paper supply, causing the printer to shut down. If this does happen, the results can be re-printed only up until the time when the sectors containing that data are "cleared" as described above in Section 4.9. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 6.2 An electronic copy of the data will be sent at the same time it is printed to the Windows-based computer, and stored in the following folder: D:\Dawning Files\results. The files should be transferred as soon as the run is finished into the following folder within that one: D:\Dawning Files\results\H L B S\[test]\[set#], where [test] equals "HDL Cholesterol," "Cholesterol & Glucose," or "Triglycerides," and [set#] equals "4th set," "5th set," etc. One folder should be used for each "run" where a "run" is defined as the maximum number of files for which tests were run sequentially and did not re-use the same numbered sector. Thus, once an individual sector is re-used for more samples (whether they are the same test or not), those files should be stored in a new folder as part of a different "run." One the run is completed, its files should be condensed into a Zip file labeled "4th_set.zip," "5th_set.zip," etc. A copy of each zip file should be stored both in the folder labeled "zipped files" and on a 3.5"disk to be transferred to the data analyst. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 7.0 | Safety: No known safety concerns. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 8.0 | Time and capacity: (The following is a rough estimate of time required for one technician to process 80 samples, although the work is much more efficient if several people work together for some of the tasks.)
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 9.0 | Protocols: Mutagenized mice are bled for these chemistries at 8 weeks of age after consuming a regular chow diet, and again at 13 weeks of age after consuming a high fat, high cholesterol "atherogenic" diet. | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| 10.0 | Notes: Be sure to plan the sequence and timing of all tasks carefully. There are numerous procedures within this protocol that may not be stopped in the middle, such as the HDL Cholesterol sample preparation, followed immediately by the HDL test. The reagent cartridges in the CX5 need to be changed regularly which interrupts the run. The cartridge needs to be changed, calibrations done, controls run, and then the run of samples can be resumed. Plan well and good luck! |